Assessment of the measurement of canine and feline serum fibroblast growth factor-23 concentrations by automated chemiluminescence immunoassay.
- Journal:
- The Journal of veterinary medical science
- Year:
- 2022
- Authors:
- Miyakawa, Hirosumi et al.
- Affiliation:
- Nippon Veterinary and Life Science University · Japan
Abstract
This study compared canine and feline fibroblast growth factor (FGF)-23 concentration measurements between automated chemiluminescence assay (CLEIA) and enzyme-linked immunosorbent assay (ELISA). Seventy serum samples each from dogs and cats were evaluated. FGF-23 measurements by CLEIA significantly correlated with those of ELISA in both dogs and cats. The Bland-Altman test showed that FGF-23 between CLEIA and ELISA had fixed and proportional biases, respectively, in both dogs and cats. Measurements by CLEIA were lower than those of ELISA, especially in higher serum FGF-23 concentrations. This study showed that FGF-23 concentrations in dogs and cats can be evaluated by automated CLEIA. However, FGF-23 cannot be directly compared between CLEIA and ELISA.