Compact breast shape acquisition system for improving diffuse optical tomography image reconstructions.
- Year:
- 2023
- Authors:
- Vanegas M et al.
- Affiliation:
- Department of Bioengineering · United States
Abstract
Diffuse optical tomography (DOT) has been investigated for diagnosing malignant breast lesions, but its accuracy relies on model-based image reconstructions, which in turn depends on the accuracy of breast shape acquisition. In this work, we have developed a dual-camera structured light imaging (SLI) breast shape acquisition system tailored for a mammography-like compression setting. Illumination pattern intensity is dynamically adjusted to account for skin tone differences, while thickness-informed pattern masking reduces artifacts due to specular reflections. This compact system is affixed to a rigid mount that can be installed into existing mammography or parallel-plate DOT systems without the need for camera-projector re-calibration. Our SLI system produces sub-millimeter resolution with a mean surface error of 0.26 mm. This breast shape acquisition system results in more accurate surface recovery, with an average 1.6-fold reduction in surface estimation errors over a reference method via contour extrusion. Such improvement translates to 25% to 50% reduction in mean squared error in the recovered absorption coefficient for a series of simulated tumors 1-2 cm below the skin.