Peer-reviewed veterinary case report
Evaluation of agreement between invasive and non-invasive blood pressure measurement using the PetMAPdevice in rabbits.
- Journal:
- Frontiers in veterinary science
- Year:
- 2023
- Authors:
- Sredenšek, Jerneja et al.
- Affiliation:
- University Clinic for Companion Animals
- Species:
- rabbit
Abstract
BACKGROUND: Measurement of arterial blood pressure is recommended in anaesthetized animals to guide perioperative treatment. Invasive blood pressure measurement is considered the gold standard, however it is also technically challenging, requires specialised equipment and carries certain risks. For these reasons, non-invasive blood pressure measurement devices are commonly used and are expected to provide accurate and reliable results. This requirement is particularly true for rabbits, in whom peri-anaesthetic hypotension is commonly observed and in whom perioperative mortality remains disproportionally high. Several authors have compared different non-invasive devices with invasive measurements in rabbits and have reported contrasting results. However, to date no comparison between invasive measurements and the PetMAPdevice, that has been designed specifically for veterinary medicine, has been reported. AIM AND HYPOTHESIS: The aim of the study was the comparison of invasive blood pressure measurement with PetMAPin rabbits. We hypothesised that PetMAPwould show acceptable agreement with the invasive measurements according to the American College of Veterinary Internal Medicine guidelines. MATERIALS AND METHODS: Sixteen client-owned rabbits presenting for various surgical interventions were included in the study. Invasive measurements were performed by cannulation of an auricular artery. The PetMAPcuff was applied distal to the elbow according to the manufacturer's guidelines. For each measurement with PetMAP, three invasive blood pressure values were recorded. The mean of the three invasive values was compared with one value measured with PetMAP. RESULTS: Data collected from 16 rabbits were used for statistical analysis. In the clinical setting, the PetMAPdevice showed significant overestimation of systolic, diastolic and mean arterial pressure, which were measured in the auricular artery. In addition, the bias was not constant, implying that the device poorly predicted changes in blood pressure. CONCLUSION: The PetMAPdevice did not meet any of the American College of Veterinary Internal Medicine recommendations.
Find similar cases for your pet
PetCaseFinder finds other peer-reviewed reports of pets with the same symptoms, plus a plain-English summary of what was tried across them.
Search related cases →Original publication: https://pubmed.ncbi.nlm.nih.gov/37492435/