Ex vivo comparison of different thoracoabdominal stapler sizes for typhlectomy in canine cadavers.
- Journal:
- Veterinary surgery : VS
- Year:
- 2022
- Authors:
- Matz, Brad M et al.
- Affiliation:
- Department of Clinical Sciences · United States
- Species:
- dog
Abstract
OBJECTIVE: To determine the influence of staple size on leakage pressure of typhlectomy sites in canine cadavers. STUDY DESIGN: Randomized, experimental cadaveric study. ANIMALS: Twenty-four fresh canine cadavers. METHODS: Ileocecocolic segments were exteriorized following right paracostal laparotomy after euthanasia. Cecal base length and wall thickness were measured. Each cecum was randomly assigned to 1 of 3 groups (TA 30 V3 2.5 mm, TA 60 3.5 mm, and TA 60 4.8 mm). The cecal base was stapled and the cecum was removed. A 10 cm segment including the stapled cecal excision site was tested for initial leak pressure. RESULTS: The mean ± standard deviation body weights across the groups were 18.7 ± 6.1 kg, 16.2 ± 7.5 kg, and 14.2 ± 5.5 kg for the TA 30 V3 2.5 mm, TA 60 3.5 mm, and TA 60 4.8 mm groups, respectively (P = .48). There were no differences for mean cecal base length or wall thickness. Mean initial leak pressure (ILP) across groups was 182 ± 111 mmHg (TA 30 V3 2.5 mm), 112 ± 57 mmHg (TA 60 3.5 mm), and 77 ± 60 mmHg (TA 60 4.8 mm) (P = .78). CONCLUSION: Each stapler size that was evaluated resulted in a mean ILP in excess of typical intraluminal pressures under normal circumstances. There were no differences among groups. CLINICAL SIGNIFICANCE: The results of this cadaveric study support the use of any of the stapler sizes evaluated in similarly sized dogs. A prospective study is needed to be able to correlate stapler size and clinical outcome.
Find similar cases for your pet
PetCaseFinder finds other peer-reviewed reports of pets with the same symptoms, plus a plain-English summary of what was tried across them.
Search related cases →Original publication: https://pubmed.ncbi.nlm.nih.gov/35191557/