In vitro evaluation of a novel prosthesis for laryngoplasty of horses with recurrent laryngeal neuropathy.

Journal:
Equine veterinary journal
Year:
2000
Authors:
Schumacher, J et al.
Affiliation:
Department of Farm Animal and Equine Medicine and Surgery · United Kingdom
Species:
horse

Abstract

A prosthesis, composed of a steel cable and stress-reducing washers, was developed to prevent failure of laryngoplasty, a common treatment for horses affected by recurrent laryngeal neuropathy. Laryngoplasties were performed on 15 cadaveric larynges using a polyester suture on one side and the cable prosthesis on the other. Each prosthesis was distracted at a displacement rate of 20 mm/s using a servohydraulic materials testing machine until laryngoplasty failed. Distraction force and actuator displacement were recorded and analysed. All 15 laryngoplasties performed with a suture failed at the muscular process at a mean +/- s.d. force of 55.8 +/- 13.1 N. Six laryngoplasties performed with the cable prosthesis failed at the muscular process at mean force 219.6 +/- 125.0 N. In the other 9, the arytenoid cartilage was avulsed from the larynx at mean force 206.4 +/- 75.3 N, and the cable then tore through the muscular process at mean force 357.0 +/- 32.0 N. The difference in force required to cause failure of laryngoplasty was significant (P<0.0001). Although the prosthesis resisted substantially higher forces than did the suture, the effects of the prosthesis in vivo must be evaluated.

Find similar cases for your pet

PetCaseFinder finds other peer-reviewed reports of pets with the same symptoms, plus a plain-English summary of what was tried across them.

Search related cases →

Original publication: https://pubmed.ncbi.nlm.nih.gov/10661384/